ULTRASONIC COATING
SYSTEM-UCS2000
EQUIPMENT
OVERVIEW
UCS2000 is an ultrasonic spraying equipment used for drug balloon and stent. The equipment uses high-frequency ultrasonic nozzles and an integrated precision injection pump, which can achieve smaller drug atomization particle size and improve the uniformity and consistency of stents. UCS2000 is equipped with precise spindle that runs smoothly and holding chuck with three internal needle finges stuiable for stents of various sizes and materials . In addition, UCS2000 adopts a special catheter looping mounting system allowing any catheter length, any luer or no luer at all of drug balloon catheters, fundamentally solving the problem of holding ultra long catheters.
Firgure 1 UCS2000 Ultrasonic Coating System
Ultrasonic atomization system |
Ultrasonic nozzle frequency |
130KHz |
Ultrasonic power |
0-5W |
|
Ultrasonic power control accuracy | 0.1W |
|
Spray flow |
0.01-2ml/min |
|
Diversion pressure |
0-15psi |
|
Can spray balloon length |
0-300mm |
|
Injection pump system |
Syringe size | 10μl-60ml Optional |
Flow range | 0.01μl/min-43.35ml/min |
|
Motor system |
Accuracy of linear motion | 0.01mm |
Linear maximum velocity | 30mm/s |
|
Maximum rotational speed | 300r/min |
TECHNICAL
PARAMETER
- High-frequency ultrasonic nozzle
- Special catheter looping mounting system allowing quick catheter loading/unloading
- Three internal needle finges stuiable for stents of various sizes and materials
- Integrated precision injection pump
PERFORMANCE
CHARATERISTICS
Figure 5 Spray process Settings page
Innova Design, Inc .
Tel: (001) 858-535-9389 (US)
(0086) 0371-60986806 18603720011 15603712688 (China)
Web:www.innovadesign.com
E-mail: info@innovadesigninc.com(US) info@zbamed.com (China)
Add: 13230 Evening Creek Dr. S, Suite 216, San Diego, CA 92128, USA
Figure 3 Ventilation system
Ultrasonic Nozzle and Syringe Pump
- Ultrasound nozzle with 130 KHz frequencies
- Ultrasound power 0-5W with accuracy 0.1W
- Flow rate 0.5-2mm
- Ultrasound syringe pump for suspension solutions
Ventilation System
- High volume ventilation system allowing excessive fume to be expelled
- Removable ventilation filter allowing quick cleaning
- External high pressure fan, easy to clean and replace
Catheter holding fixture
- Special catheter looping mounting system, allowing any catheter length, any luer or no luer at all
- Spindle speed 0-300 Rpm
- Hollow shaft spindle with large opening and guiding clamp for quick catheter loading/unloading
Control System
- 12” Touch screen controller, Support English and Chinese language
- Allow 1000 recipes stored in HMI
- Real time display of machine status and parameter values
Figure 2 Ultrasound syringe
Figure 4 Control system interface
Figure 5 Coating process interface